Automated MRI quality report

volume t0145 · T1-weighted brain, 3-D · agent model openai/gpt-6.1-terra

the original scan
the original scan
the final scan
the final scan
final − original (red: brighter now)
final − original (red: brighter now)

Findings

artifactagent's answerrepair keptchecked on the final scanthe agent's note
Noise (grain)foundstep 2: denoisingclear
Spike (stripes)foundstep 1: spike removalclear
Ghosting (faint copies of the head)not found–clear
Bias (uneven brightness)foundstep 3: bias correctionclear
Motion (head movement)uncertainno repair exists: the rescan decisionclearAfter repairs, the final scan reading was 0.000 with no significant front–back leak.

What the agent did

  1. Spike removal
  2. Denoising
  3. Bias correction

14 measurements, 3 looks at the images, 3 repairs tried, 3 kept.

Impression

Noise, spike stripes and shading were substantially reduced. No ghosting or definite motion was detected, though motion cannot be excluded.

Recommendation

Rescan: no

Post-repair motion estimate 0.000 is below the 0.033 rescan level, with no significant front–back leak; under-reading remains possible.

Return the final scan for clinical review alongside the original. No motion-based rescan is indicated by the available measurement; rescan if diagnostic review finds image quality inadequate.

Checked against the run log

✓ the repairs it lists are the ones applied · ✓ every number it quotes was measured · ✓ its yes/no answers match its findings

The answers, notes, impression and recommendation are the agent's words, shortened, with version labels and function names replaced by what had been done. The repair kept and the check on the final scan come from the run log.